Permanent vibration monitoring for systems such as motor, pump and blower.
Monitoring system suitable for multi-channel measurement.
Frequency range, measurement range or output signal can be customized.
Features
- ■ Mixed implementation with the conventional system “VM-9201” is possible
- ■ A wide variety of pickups
-
Selectable from electrodynamic velocity pickup or piezoelectric accelerometer.
- Electrodynamic velocity pickup
-
Specialized in medium frequency (up to 1,000 Hz) vibration detection. Correspond to velocity monitor.
Type VP-3144 C/D VP-3354 A VP-3364 A VP-3134 A VP-3213 AC/AD VP-3133 H/V Sensitivity 10mV/(mm/s) 10mV/(mm/s) 10mV/(mm/s) 10mV/(mm/s) 17.5mV/(mm/s) 17.5mV/(mm/s) Natural
frequency14Hz 14Hz 14Hz 14Hz 4.5Hz 4.5Hz Operational
temperature
range-20 to +80 °C -20 to +80 °C -20 to +80 °C -10 to +70 °C -20 to +70 °C -10 to +70 °C Structure Drip-proof Water-proof Water-proof Flame-proof Drip-proof Flame-proof Outward
appearanceType C
Type AC
Notes High sensitivity/
medium frequencyMedium frequency /
2-directionalMedium frequency /
3-directionalMedium frequency Low frequency /
horizontal onlyLow frequency /
H (horizontal only),
V (vertical only) - Piezoelectric accelerometer
-
Specialized in high frequency (over 1,000 Hz) vibration detection. Correspond to acceleration monitor.
Type VP-A51 IW VP-4135 VP-A4345Ⅰ Sensitivity 5mV/(m/s2) 5mV/(m/s2) 5mV/(m/s2) ±10% Natural frequency 30,000 Hz 8,000 Hz 25,000 Hz Operational temperature range -30 to +110 °C -10 to +60 °C -20 to +80 °C Structure Water-proof Flame-proof Safe and explosion proof ⅡBT4 Outward
appearanceNotes Insulated ― ―
- ■ Multi-channel
- Maximum 12 channel per 1 system is available. Range and function can be set individually per every channel.
Application examples
Cable is drawed from pickups located at large sized motor at hope power generator to a monitoring system inside of the control panel. Monitoring systems can monitor vibration values and send out the alarm in an emergency.
Specifications
Type | VM-9301 series | VM-9301A series | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Pickup | Electrodynamic velocity pickup | Piezoelectric accelerometer | |||||||||
Amplifier unit type |
|
|
|||||||||
Frequency range |
Low: 5 to 500Hz (-11 to +6%) Medium:10 to 500Hz (-11 to +6%) Filter selectable (pass band)
|
5 to 10kHz (-30 to +6%) Filter selectable (pass band)
(High-pass filter : 20Hz- in case of displacement |
|||||||||
Measurement range |
Displacement | 50, 100, 150, 200, 300, 500, 999 (1,000) μmP-P | 100, 150, 200, 300, 500, 999 (1,000) μmP-P | ||||||||
Velocity | 10, 15, 20, 30, 50 mm/s | 10, 15, 20, 30, 50 mm/s | |||||||||
Acceleration | – | 10, 15, 20, 30, 50, 100, 150, 200, 300, 500 m/s2 | |||||||||
Equipment failure |
Indication lamp |
Orange LED located on the upper part of amplifier front panel [FAT] Secondary power is lit at ubnormal condition. | |||||||||
Alarm contact |
1a contact (open at normal condition) operate under fuse melting, power switch shut off and primary / secondary power failure | ||||||||||
Alarm reset | Select from automatic or manual recovery | ||||||||||
Pickup cable break alarm |
Indication lamp |
Green LED located on the upper part of amplifier front panel [PU]
Lights-out when wire is broken (lights-on at normal condition) (When pickup connected with charge amplifier is used, this function is not valid.) |
|||||||||
Alarm contact |
Standard:1a contact *available 1b or 1c (open at normal condition, closed when wire is broken) operate when pickup cable is broken. | ||||||||||
Alarm reset | Select from automatic or manual recovery | ||||||||||
Alarm | Alarm step | Type 1 line: upper limit 2 steps Alarm (ALM) / Trip (TRP) Type 2 line: upper limit 1 step Alarm1 (ALM1,1lline), Alarm2(ALM2, 2line) |
|||||||||
Delay timer | Select from 0 to 15 sec. by the slide switch inside of the amplifier unit (0 sec. if not specified.) | ||||||||||
Setting extent |
5.0 to 99.9% (can be set at 0.1% intervals) (When alarm function is not used 99.9% is set. Indication lamp and contact is on at over full scale value) |
||||||||||
Indication lamp |
Type 1 line: Orange LED [ALM], red LED [TRP] Type 2 line: Orange LED [ALM1], red LED [ALM2] | ||||||||||
Alarm Contact |
Select from “ALM / ALM1 circuit:2a contact TRP / ALM2 circuit 2a contact” or “ALM / ALM1 circuit:1b1a contact TRP / ALM2 circuit 1a1b contact” |
||||||||||
Contact capacity | AC110 V 2 A (DSP Relay made by Panasonic) DC30 V 0.2 A | ||||||||||
Alarm reset | Select from automatic (Hysteresis 3%) or manual recovery | ||||||||||
Rectification level output |
DC output 1 |
Select from DC4-20mA (insulation output), Loading 500Ω or less or DC1-5 V (insulation output), Loading 100kΩ or more |
|||||||||
DC output 2 |
Select from DC4-20mA (insulation output), Loading 500Ω or less or DC1-5 V (insulation output), Loading 100kΩ or more |
||||||||||
Waveform output |
Pickup waveform output |
AC±10mV /(mm/s) or AC±17.5mV /(mm/s), Loading 100kΩ or more (Depending on specifications of pickup) |
AC±5mV/(m/s2), Loading 100kΩ or more (Depending on specifications of pickup) |
||||||||
Waveform output |
AC±5V(at the full-scale), Loading 100kΩ or more | ||||||||||
Panel meter | Red LED three-digit indication, [9.9.9] is displayed at over 120% of full-scale | ||||||||||
Operational temperature range |
0 to 50 °C, 20 to 95%RH, wet bulb temperature is 35°C (non-condensing) | ||||||||||
Power supply | Available range AC85 to 264V, 47 to 66Hz | ||||||||||
Power consumption |
20VA or less (1CH), 30VA or less (3CH), 45VA or less (6CH), 60VA or less (9CH), 75VA or less (12CH) At AC100V (Depends on the number of amplifier or power supply) |
||||||||||
Painting color | Munsell 5Y7/1, half matted | ||||||||||
Mass | Approx. 2.0kg (1CH), approx. 3.9kg (3CH), approx. 5.9kg (6CH), approx. 8.2kg (9CH), approx. 10.2kg (12CH) |
External dimensions
Unit:mm
A | B | C | D | E | F | G | H | I | |
---|---|---|---|---|---|---|---|---|---|
Type 1 channel | 90 | 170 | 300 | 88 | 132 | 90 | 148 | 162 | 50 |
Type 3 channel | 210 | 149 | 300 | 160 | 132 | 180 | 143 | 100 | 195 |
Type 6 channel | 300 | 149 | 300 | 250 | 132 | 270 | 143 | 100 | 285 |
Type 9 channel | 390 | 149 | 300 | 340 | 132 | 360 | 143 | 100 | 375 |
Type 12 channel | 480 | 149 | 300 | 430 | 132 | 450 | 143 | 100 | 465 |